11
Cause
The measured temperature is below 95°F (35°C)
or above 107.6°F (42°C).
Instructions
1. Check correlation between the selected mode icon
and the actual measurement location used.
2. Turn the unit OFF and then ON again.
3. Retake temperature using the proper technique
for oral, axillary, and rectal method.
4. Any of the methods may be used for accuracy
depending on individual circumstances.
Cause 1
The thermometer is too cold or too hot to operate.
Operating temperature range is 50-104°F (10-40°C)
Cause 2
The thermometer was placed in the measurement
method before the “hourglass” symbol disappeared
and before two “Beeps” were heard.
DISPLAY MESSAGES
Displayed
The thermometer has been designed to display feedback messages
to ensure proper usage and accurate measurement. In the event
of a malfunction or incorrect measurement, the following
messages will be displayed:
Instructions
1. Bring the thermometer to a room where the temperature is between 50°F
and 104°F (10°C and 40°C). Thermometer should be allowed to stabilize
at room temperature for at least 45 minutes.
2. Wait at least one minute and then turn the thermometer ON again.
3. Wait for the “hourglass” symbol to disappear and two “beeps”
to be heard.
4. Retake temperature using the proper technique. (as described on page 9)