1INSTRUCTION MANUALOMRONDigital Automatic Blood Pressure Monitor ModelHEM-907HIDEP-SET100AUTO140180220260280MODEAVG.MANU.CHECKSINGLEON/OFFAVG./1st/2nd
10Notes on SafetyCautionBe sure to use the power supply of 230 VAC.It may cause re or electric shock.Do not share an electric outlet with other unit
11Notes on SafetyMaintenanceBe sure to inspect the unit on regular basis. Check accuracy on a 1 year interval.If the unit has not been used for a whil
12Components of the ProductMain unit AccessoriesHIDEP-SET100AUTO140180220260280MODEAVG.MANU.CHECKSINGLEON/OFFAVG./1st/2nd/3rdDEFLATIONSTARTSTOPMedium
13Select the arm cuff according to the arm circumference of the patient.Arm circumference Name of the arm cuff17 - 22 cm22 - 32 cm32 - 42 cmSmall size
14How to Apply the Arm Cuff (cont.)RANGEART.RANGEART.INDEX22~32cm(9"~13")REFHEM-CR19Adjust the level of the arm cuff to the level of the hea
15How to Measure Blood PressureWarningIf any abnormal matter occurs during the mea-surement such as the in ation does not stop, remove the arm cuff o
16How to Measure Blood Pressure (in SINGLE Mode) Wrap the arm cuff over the patient’s arm.1.Set the MODE Selector to “SINGLE”.3.ON/OFFAVG./1st/2nd/3rd
17Wrap the arm cuff over the patient’s arm.1.Set the MODE Selector to “AVG.”.3.ON/OFFAVG./1st/2nd/3rdDEFLATIONSTARTSTOPHIDEP-SET100AUTO140 18022026028
18How to Measure Blood Pressure (in MANU. Mode) Wrap the arm cuff over the patient’s arm.1.Push the ON/OFF (power) Button to turn on the power.3.ON/OF
19How to Measure Blood Pressure (in CHECK Mode)Display of pressure can be con rmed by the CHECK Mode (yearly procedure).What you need to prepare(1) W
2Intended UseMedical Purpose This is a medical device that measures the brachial blood pressure in a non-invasive manner.Intended User Legally cert
20Installation and Replacement of Battery PackRemove the battery cover on the back of the main unit.Installation: To install the battery pack, connect
21Clean the blood pressure monitor,the cuff and the cables using a cloth dampened with water,diluted disinfectant alcohol or diluted mild detergent.To
22List of Error CodesError code Cause How to correctEr1In ation errorWhen the pressure does not exceed 15 mmHg after operating a pump for 15 seconds
23If a trouble occurs while using the unit, please check the following.Problem What to inspect How to correctThe unit in ates to abnormally high (low
24Speci cationsName : OMRON Digital Automatic Blood Pressure Monitor HEM-907Model : HEM-907-E / HEM-907-E7Display : Digital displayMeasurement : Os
25 Manufacturer OMRON HEALTHCARE Co., Ltd.53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPANEU-representative OMRON HEALTHCARE EUROPE B.V.Scorpius
26EnglishImportant information regarding Electro Magnetic Compatibility (EMC)With the increased number of electronic devices such as PC.s and mobile (
3ExemptionsOMRON will not bear any responsibilities on the following matters.When a problem or damage occurs caused by the maintenance and/or repair c
4Names and Functions of the PartsHIDEP-SET100AUTO140 180220260280MODEAVG.MANU.CHECKSINGLEON/OFFAVG./1st/2nd/3rdDEFLATIONSTARTSTOPmmHgSYSmmHgDIABPMPULS
5Battery levelDisplays for the usable level.Displays for the low level.Displays for the unusable level.Names and Functions of the PartsDisplayHIDE
6Names and Functions of the PartsExplanation of Functions(1) Pressure setting function P-SETIn ation can be set by AUTO (automatic setting) or the ta
7Names and Functions of the Parts(4) Setting mode change functionYou can set the number of times to measure, the waiting time until the start of measu
8Notes on SafetyThe warning signs and the sample icons shown here are listed for you to use the product safely and correctly as well as to prevent the
9Notes on SafetyCautionWhen you are not going to use the unit for a long period of time, be sure to remove the AC adapter from the electric outlet.You